Zeiss 160 kVp Versa 610

This scanner uses a two-stage magnification approach to achieve sub-micron spatial resolution at source to object distance much greater than these allowed in conventional XCT systems. Zeiss’s resolution at a distance, RaaD™, approach combines the geometric magnification of the X-Ray imaging with a smart system of microscope lenses to further magnify the image. The system is fitted with X0.4, X4, X20, and X40 magnification lens mounted on a barrel mount for rapid change over. This means that the operator can achieve submicron resolution (pixel size ≥ 70 nm or spatial resolution ≥ 0.5 µm) at large working distances, allowing for imaging of larger specimens or small specimens in contained in-situ rings. Importantly, the machine is capable of phase-contrast imaging overcoming an important limitation of convectional absorption-based CT imaging; i.e. low or complete absence of contrast for low Z materials. The latter is particular useful for biomedical imaging, CFRP and polymer imaging applications.

As of Spring 2025 our Zeiss 160kVp Versa 510 has been upgraded into a 610.  The key upgrades include:

  • Increased power from 10W to 25W
  • More efficient objective optics
  • Upgraded higher specification PC

These upgrades mean the same image quality can be achieved with reduced scan times, this enables higher throughput scanning, and will make scans that were previously not practical to complete in a timely manner can now be considered.

location

Highfield Campus

Summary

‣ 30 -160 kVp
‣ 25W
‣ 2048 x 2048 pixel detector
‣ multiple magnification objectives up to 40x
‣ 0.5 µm true spatial resolution
‣ phase-contrast imaging mode